WHY CHOOSE Easypharma Ga/Ge hot cell

  • Limited weight and footprint, allowing perfect and easy integration into all departments
  • Cleaning is simple and fast thanks to the fully opening front door
DESCRIPTION
CHARACTERISTICS
ADDITIONNAL PRODUCT

DESCRIPTION

The Easypharma Ga/Ge manual shielded hot cell with 2 glove ports, is designed for radiolabelling and fractionating of radiopharmaceuticals mainly labelled with Gallium-68 for PET activity. It allows the elution of the Germanium generator as well as radiosynthesis operations with modules adapted to the workspace if necessary.

Its compact design, hides its versatility, since it also allows fractionating of ready-to-use radiopharmaceuticals such as 18F- DG or any other fluorinated drug with its optimised architecture integrating a dedicated and secure space, equipped with a shielded pot lift.

Easypharma Ga/Ge: compact and versatile

The Easypharma Ga/Ge cabinet consists of:

  • The 316L stainless steel work surface with a special acid-resistant coating is easily decontaminated and provides the minimum pace required to perform routine Gallium-68 operations and 18F-FDG backups. It is equipped with 3 independent accesses: one for the dose calibrator, one for the waste bin and one for the pot lift; and with two cable connectors giving acess to the Germanium generator located in the shielded housing below. Electrical, USB or RJ45 sockets are available as well as LED lighting above the work surface, around the periphery of the filter, and two UV germicidal tubes with timers, on either side.
  • The 40 mm lead shielded dose calibrator compartment allows the installation of the measuring device of choice for enhanced versatility. The optional “Posilift” pneumatic dipper raising system allows to reduce operator exposure and provides unparalleled freedom of movement, access to the dose calibrator chamber and improved ergonomics.
  • The lockable generator/pot compartment with a 40 mm lead shield, for improved security, allows both the positioning of the Germanium generator and is equipped with an electric lift compatible with several shielded carrier pots for multi-dose vials of radiopharmaceuticals.
  • The waste compartment with a 40 mm lead shield, is accessible from the work surface and guarantees protection, even in the eventof decay of a vial with resi dual activity. Partial opening of the magnetic cap significantly reduces exposure to the hands and fingers during handling operations in the workspace.
  • The lateral airlock chambers with a 20 mm lead shield and viewing window, are each equipped with a robust stainless steel sliding tray, to remove and insert equipment and patient doses. For improved security, the inner and outer door openings are interlocked with foot-operated control for the inner door.

Biological protection tailored to your needs

It provides the operators with daily biological protection adapted to their needs, consisting of 40 mm lead screens and 120 mm Diamond Glass scratch and chemical resistant laminated lead glass. The integrity of the preparations is guaranteed by a Class A (ISO 5) laminar airflow over the entire workspace.

Learn more about the available options

Easypharma Ga/Ge complies with the standards 2006/42/EC, 2004/108/EC, 96/29 EURATOM, ASNR resolution no. 2014-DC-0463, as well as ISO 14644.

CHARACTERISTICS

  • External dimensions (1/2 airlock chambers): L 1,223/1,610 x D 838 x H 2,240 mm
  • Hot cell tare weight (1/2 airlock chamber): 2,890/2,990 kg
  • Exterior finish: ABS
  • Interior finish: PVC
  • Standard equipment:
    - 1 airlock chamber
    - 1 shielded compartment for generator and pot with lift
    - 1 shielded compartment - dose calibrator
    - 1 shielded compartment - 1 waste bin
    - shielded cable chicane (max. 9 passages)
  • Type of lighting: LED
  • Brightness: > 1,000 Lux (ajustable via remote control)
  • Germicidal UV: 1 UV-C 8W tube with timer
  • Noise level: < 63 dB(A)
  • Shielding thickness: 40 mm of lead

ADDITIONNAL PRODUCT